| 1 | +# Mirion Lynx2 HPGe detector |
| 2 | + |
| 3 | +The Lynx HPGe detector is a gamma detector, used on VESUVIO as a gamma detector. It may also be called |
| 4 | +a "digital signal analyzer" or a "multichannel analyzer", and is made by a company called Mirion. |
| 5 | + |
| 6 | +## Manual |
| 7 | + |
| 8 | +A manual for the device is available at: |
| 9 | +``` |
| 10 | +\\isis\shares\ISIS_Experiment_Controls\Manuals\lynx2-gamma-detector |
| 11 | +``` |
| 12 | + |
| 13 | +## Hardware / Networking |
| 14 | + |
| 15 | +We do **not** talk to this device from IBEX in any way. |
| 16 | + |
| 17 | +The hardware is an ethernet device, which should be set up in DHCP mode to automatically acquire an IP address. |
| 18 | + |
| 19 | +Configuration screens for the hardware are available by visiting its IP address in a web browser. Use: |
| 20 | + |
| 21 | +``` |
| 22 | +http://x.x.x.x/mainex.asp |
| 23 | +``` |
| 24 | + |
| 25 | +to access the configuration screens, as the _home_ page requires flash player, but the hardware configuration |
| 26 | +screens do not. |
| 27 | + |
| 28 | +If the device is (incorrectly) set to static IP mode, it will likely default to an IP of `10.0.0.3`. If the device has |
| 29 | +forgotten its DHCP setting, then you may need to connect to the device directly (point-to-point) from NDCVESUVIO or |
| 30 | +another computer, with netmask `255.255.255.0`. |
| 31 | + |
| 32 | +The settings *should* look like this: |
| 33 | + |
| 34 | + |
| 35 | + |
| 36 | +## Software |
| 37 | + |
| 38 | +The device is controlled using a program called "GENIE2K", which runs on **NDC**VESUVIO. This software is licensed to |
| 39 | +a specific computer, so cannot be run on another computer. This software is unrelated to OpenGENIE - it is manufacturer |
| 40 | +software. It does not integrate with IBEX in any way. |