Skip to content

RSNA-QIBA-US-SWS/QIBA-DigitalPhantoms

Folders and files

NameName
Last commit message
Last commit date

Latest commit

 

History

89 Commits
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Repository files navigation

DOI License: MIT

QIBA-DigitalPhantoms

RSNA QIBA wiki overview of US SWS Digital Phantom Studies

Related publication to cite: Palmeri ML, Qiang B, Chen S, Urban M. "Guidelines for Finite-Element Modeling of Acoustic Radiation Force-Induced Shear Wave Propagation in Tissue-Mimicking Media," IEEE TUFFC, 64(1):78-92, 2016.

This repository contains the parameters and simulation configuration files used for the digital phantom datasets that will be circulated to all of the manufacturers.

Curvilinear Probe Parameters

  • Radius of curvature: 60 mm
  • Element Height: 14 mm
  • Element Pitch: 0.477 mm (0.007 mm kerf)
  • Center Freq: 3.0 MHz
  • Frac. Bandwidth: 100%
  • Elevation Focus: 50 mm

ARF Excitation Parameters

  • Frequency: 3.0 MHz
  • F/# = 2, 3.5
  • 500, 1000 cycles
  • Focal Depths: 30, 50, 70 mm

Acoustic Material Properties

  • Attenuation: 0.45 dB/cm/MHz
  • Linear

Elastic Material Properties

  • Poisson’s ratio: 0.495
  • Shear modulus: [1.0, 2.0, 5.0, 10.0] kPa
  • LS-DYNA Input Deck: dyna/qiba_elastic_pml.dyn
  • LS-DYNA Run Script: dyna/qiba_elastic_pml.py
  • ABAQUS Input Files: abaqus/

Viscoelastic Material Properties

  • G_0 = 10 kPa, G_infinity = 2 kPa, beta = 6666.7 Pa-s
  • G_0 = 15 kPa, G_infinity = 4 kPa, beta = 5500.0 Pa-s
  • G_0 = 20 kPa, G_infinity = 4 kPa, beta = 4000.0 Pa-s
  • LS-DYNA Input Deck: dyna/qiba_ve_pml.dyn
  • LS-DYNA Run Script: dyna/qiba_ve_pml.py
  • FEBio Input File: febio/phantom1_200Hz_case1.feb

Misc. Parameters

  • Displacement data in the imaging plane saved at 0.1 mm spacing every 0.1 ms.
  • 15 mm of total lateral extent
  • Just axial components of displacements
  • Just imaging plane displacements
  • MATLAB-formatted data will be shared

Mesh Dimensions & Configurations

  • ¼ symmetry
  • 0.25 mm node spacing
  • 10 mm elevation
  • 25 mm laterally
  • Focal depth + 25 mm axially
  • PMLs on outer lateral/elevation faces and top/bottom

Future Efforts

  • Ultrasonic tracking configurations and realizations

Download Digital Phantom Data

The raw simulation data for processing is available through the Duke University Library Digital Research Data Repository: QIBA Ultrasound Shear Wave Speed Digital Phantoms

Contributors

  • Dr. Shigao Chen, Ph.D.
  • Dr. Mark Palmeri, M.D., Ph.D.
  • Bo Qiang, MSEE
  • Dr. Matthew Urban, Ph.D.
  • Dr. Jingfeng Jiang, Ph.D.

Funding

This work has been supported in part by RSNA QIBA, FDA contracts HHSF223201400703P and HHSN268201300071C, and NIH grants R01DK092255 and R01 EB002132.

About

RSNA QIBA Digital Phantom Studies

Resources

License

Stars

Watchers

Forks

Packages

No packages published

Contributors 2

  •  
  •